TECHNICAL SPECIFICATIONS : Intensity (220-240V)

Weight and Dimensions
Height : 4.5 inch ( 11.5 cm)
Width : 14.4 inch (36 cm)
Length : 17.8 inch (45 cm)
Weight : 12 Kg.

Low Frequency Leakage (50-60Hz)
Patient leakage < 150 µArms

Cooling Single speed fan

High Frequency Leakage
Less than 150 mA rms., All output

 

Input Power Requirements
Volts : 170-270 VAC
Amps : 6A
Input : 1.5 KW, 50/60 Hz

Audio Volume
The mode indicator tones are adjustable to a minimum level of 45 dba at 1 meter.

Operating Mode
int 10s/30s

Certified to IEC - 60601-2-2
and IS-7583
Protection Class-1. UnitType-CF.
Output- All floating Output.
CE Certified

     
(Top : Accessories
Right : Mobile Cart)
 
OUTPUT CHARACTERISTICS
  MODE OUTPUT
FREQ. (KHz)
MAXIMUM
P-P VOLTAGE
RATED
LOAD (ohms)
MAX. POWER
(WATTS)
CREST
FACTOR*
PER Date
Cycle
BIPOLAR Micro/Coag
Macro/Cut
400
400
480
480
100
100
90
95
2.5
1.5
98 N/A
N/A
MONOPOLAR
CUT
Pure/Cut 1
Blend/Cut 2
500
500
1800
2600
300
500
350
300
1.5
2.4
98 N/A
75%
MONOPOLAR
COAG
Soft/ Desiccate
Fulgurat/Fulguart
450
450
3300
9500
150
500
150
150
4.4
8.0
97 25%
65%
* Crest Factor is an indicator of a woveform's ability to coagulate bleeders without cutting effect.
* PER (Power Efficiency Rating): A measure of an electrosurgical generators ability to deliver the selected power into a wide range of impedance.
ACCESSORIES

STANDARD
D-004 : Monopolar Handle with cable

D-005 : Monopolar twin button Handel with cable (Hand switching pencil)
D-006 : Patient Plate with cable
D-006 : Bipolar forceps with cable
D-009 : Monopolar Electrode 4 Nos
D-007 : Double pedal footswitch for Monopolar

OPTIONAL
D-001 : Single Pedal Bipolar footswitch

BACK
 
Partner Sites ::-
http://ligasealer.com , http://quadrf.com , http://plasmaeffect.com , http://rfscalpel.com